Tag: AB-3000B Cellgraph

  • AB-3000B Cellgraph

    Bioluminescnece imaging system for living cell and tissues
    Cellgraph is an imaging system to detect low-level light emission in a single living cell by using a highly sensitive EM-CCD camera.

    AB-3000B CellgraphAB-3000B Cellgraph

    Purpose and Application
    • Real time reporter gene assay
    •   Transcriptional factor activity, Clock gene (chronopharmacology). Gene expression (Transfection,RNAi,etc)
    • Analysis of cell responses
    •   Drug response(Anti-cancer drug,DDS,etc),Stress response (Hormone,Inflammation,Anti-Oxidant), Cytotoxichology
    • Analysis of signal pathway
    • Signal cascade (Calcium), Apoptosis,etc
    Features
    • Acquire images of culturing cells or tissue samples on a 35 mm dish in the living state.
    • Provide a CO2 incubator equivalent atmosphere by controlling the temperature, CO2 gas concentration and humidifying.
    • Highly sensitive and accurate detection of faint bioluminescence with an EM-CCD camera.
    • Isolates and captures multicolor faint bioluminescence.
    • Easy operation via a user-friendly interface.
    • Full control from a PC with dedicated software.
    • Completely lightproof,compact design.
    Specifications
    Product Number AB-3000B
    Product Name Cellgraph
    Cooled CCD Camera CCD type: Back-illuminated EMCCD
    Active pixel: 512 x 512 pixel
    Pixel size: 16 x 16μm
    AD Resolution: 14/16 bit
    Cooling temperature:
    -80°C with an air-cooling system when the air temperature is 25°C, -90°C with a water-cooling system*
    (when the water temperature is 25°C)
    -85°C with an air-cooling system when the air temperature is 20°C, -100°C with a water-cooling system*
    (when the water temperature is 16°C)
    *The water-cooling unit is provided as an optional accessory.
    Objective Lens 4x (NA 0.53) (Other magnification lenses are also available as optional accessories)
    Stage X-Y-Z axis manual stage, Z-axis motorized stage
    Culture dish 35mm culture dish
    Constant temperature function (Room temperature +5°C) to 45°C, 0.1°C step
    Lighting White LED with dimming function Blue LED with dimming function
    Optical Filter Up to three filters can be set (525LP/560 LP/620 LP are standard equipment)
    Exposure time 30 milliseconds to 90 minutes
    Imaging interval Indefinite
    Control software OS:Windows 7/Vista/XP
    System requirement RAM: 1 GB or more (512 MB or more for XP)
    Hard Drive: 30 GB or more free disk space required
    Interface: Full size PCI slot x1, serial port x1, USB port x1
    Power (main body) AC100-240V  106VA (varies according to the components of the whole system)
    Dimensions (main body) 430 mm (W) x 600 mm (D) x 650 mm (H)
    Weight (main body) Approx. 40kg
    Optional Accessories
    CCD camera water-cooling unit Constant-temperature water circulating system, heat-resistant tube for circulation
    CO2 gas injection unit CO2 gas-mixer
    CO2 gas humidifying unit Bubbling system for gas humidification
    Perfusion culture chamber unit Chamber for perfusion
    Objective lenses 2x, 10x, 20x, 40x, 60x

    Download (PDF, 3.77MB)